To exit the Setup Mode and save changes (370):

1. Press

[ZERO] to begin exiting Setup Mode.

DISPLAY READS Enter ~ =CAL!

2. Press

[ZERO] to bypass Calibration Mode.

DISPLAY READS Enter ~ =Stor

3. Press

[TARE] to save setup changes.

DISPLAY READS Enter ~ =End

4. Press

[TARE] to complete exit.

DISPLAY READS 0.00

To exit the Setup Mode and save changes (375):

1. Press

[ZERO] to begin exiting Setup Mode.

DISPLAY READS Enter ~ =CAL!

2. Press

[CLR] to bypass Calibration Mode.

DISPLAY READS Enter ~ =Stor

3. Press

[SAMPLE/Enter] to save setup changes.

DISPLAY READS Enter ~ =End

4. Press

[SAMPLE/Enter] to complete exit.

DISPLAY READS 0.00

To exit the Setup Mode from the view-only mode (370 and 375):

1. Press

[ZERO] to begin exiting Setup Mode.

DISPLAY READS Enter ~ =End

2. Press

[TARE] (370) or [SAMPLE/Enter] (375) to complete exit.

DISPLAY READS 0.00

To exit the Setup Mode without saving changes (370):

1. Press

[ZERO] to begin exiting Setup Mode.

DISPLAY READS Enter ~ =CAL!

2. Press

[ZERO] to bypass Calibration Mode.

DISPLAY READS Enter ~ =Stor

3. Press

[ZERO] to exit without saving changes.

DISPLAY READS Enter ~ =Undo

4. Press

[TARE] to undo changes.

DISPLAY READS Enter ~ =End

5. Press

[TARE] to complete exit.

DISPLAY READS 0.00

To exit the Setup Mode without saving changes (375):

1. Press

[ZERO] to begin exiting Setup Mode.

DISPLAY READS Enter ~ =CAL!

2. Press

[CLR] to bypass Calibration Mode.

DISPLAY READS Enter ~ =Stor