Holtgreven GSE-375 User Manual

Page 84

Advertising
background image

78



This test procedure requires that the initial analog option calibration procedure has been
completed.

To test the 0-10v output mode:

1.

Enter the Setup Mode (see Setup Mode on page 15).
Chngs Poss!
P110.-- ~ F.S.= ~ 100.00

2.

Attach the voltmeter + (red) lead to pin 3 (0-10VDC) and the - (black) lead to pin 2
(ISOLATED GND) of the Analog Output connector.

3. Key

in 62002 [SELECT].

Test ~ 0-10v
Per P176

4. Press

[SAMPLE/Enter] to set the output to 0%.

0-10v ~ 0P
0.00 VDC

5. Press

[SAMPLE/Enter] to increase the output to 25%.

0-10v ~ 25P
2.50 VDC

6. Press

[SAMPLE/Enter] to increase the output to 50%.

0-10v ~ 50P
5.00 VDC

7. Press

[SAMPLE/Enter] to increase the output to 75%.

0-10v ~ 75P
7.50 VDC

8. Press

[SAMPLE/Enter] to increase the output to 100%.

0-10v ~ 100P
10.00 VDC

To test the 0-20mA output mode:



Voltmeter readings are based on the use of a 500 ohm precision resistor. Caution! Do not exceed
500 ohms.

1.

Enter the Setup Mode (see Setup Mode on page 15).
Chngs Poss!
P110.-- ~ F.S.= ~ 100.00

2.

Attach the precision resistor to pin 5 and pin 6.

3.

Attach the voltmeter + (red) lead to pin 6 (0-20 mA) of the analog output connector.

4.

Attach the voltmeter - (black) lead to pin 5 (ISOLATED GND) of the Analog Output
connector.

5. Key

in

62003 [SELECT].

Test ~ 0-20A
Per P176

6. Press

[SAMPLE/Enter] to set the output to 0%.

0-20A ~ 0P
0.00 V

7. Press

[SAMPLE/Enter] to increase the output to 25%.

0-20A ~ 25P