Holtgreven GSE-375 User Manual

Page 81

Advertising
background image

75

C

ALIBRATION

E

RRORS

F.S. ~ TooHi

The entered calibration weight will result in an over capacity condition at full scale.
Verify that the full scale (P110) and calibration weight value are correct.

F.S. ~ TooLo

The entered calibration weight will result in a full scale input signal less than the
minimum allowed. Verify that the full scale (P110) and entered weight value are
correct.

Add ~ Load!

The calibration weight is less than 0.1% of capacity. More weight is required.

ReCal ~ ???

Repeat the cal. procedure for accuracy. This prompt appears when the calibration
weight is less than 5% of capacity, or when the A/D coarse gain is adjusted.

Entry ~ Error

An invalid entry was made.

C

OMMUNICATION

E

RRORS

Par- Er

The selected parity (P202) does not match that of the connected device.

Buf- Er

The receive buffers capacity was exceeded. This indicates a handshaking
problem. Check P204 and verify proper communication port connections.

Bit- Er

The stop bit of a received character did not occur when expected. Verify that
protocol (P200 - P204) matches that of the connected device.

TrHold

Data transmission is inhibited due to a deasserted handshake. Press [ZERO] or
[CLR] to abort transmission. Check P204.

M

ISCELLANEOUS

E

RRORS

T.X.YYYY

If catastrophic errors occur in the software, a trap error may occur and freeze the
display with address information. (X = bank number and YYYY = the address of
the trap error. Press any key five seconds after viewing message to reboot the
unit).

V

IEWING

S

ETUP

While troubleshooting it may be helpful to view the setup parameter selections. This can be done
using the procedure below (even with the internal program jumper in the "NO" position). Note
that accessing the Setup Mode in this manner will not permit parameter changes.

To view the setup parameter selections (370):

1. From the Weigh Mode, press [ZERO] + [SELECT].

DISPLAY READS Setup~Enter ~ Code!

2. Press

[TARE].

DISPLAY READS -No- ~ Chgs!~P110.-- ~ F.S.= ~ 100.0

3. Navigate the Setup Mode as described in Setup Mode on page

16.

To view the setup parameter selections (375):

1. From the Weigh Mode, press 100 [SELECT].

DISPLAY READS Setup~Enter ~ Code!

2. Press

[SAMPLE/Enter].

DISPLAY READS -No- ~ Chgs!~P110.-- ~ F.S.= ~ 100.0