To select a calibration method (370):

1. Press

[ZERO] + [SELECT] to display the calibration prompt.

2. Press

[ZERO] [PRINT] [UNITS] [TARE] to access the Calibration Mode.

3. Press

[SELECT] to scroll through the five selections.

4. Press

[TARE] to establish zero.

To select a calibration method (375):

1. Press

100 [SELECT] to display the calibration prompt.

2. Press

54321 [SAMPLE/Enter] to access the Calibration Mode.

3. Press

[SELECT] to scroll through the five selections.

4. Press

[SAMPLE/Enter] to establish zero.

F

IRST

Z

ERO

The most common zeroing procedure, First Zero is used to establish a new zero (no load)
calibration reference before proceeding to span the Model 370/375. Use this method for first-time
calibration and complete recalibration.

First Zero Calibration Method Example (370):

1. From the Weigh Mode, press [ZERO] + [SELECT].

DISPLAY READS Setup

2. Press

[ZERO] [PRINT] [UNITS] [TARE].

DISPLAY READS Fast ~ Cal~First ~ Zero? ~ -0.26

3. Remove any load on the scale.

DISPLAY READS First ~ Zero? ~ -0.42

4. Press

[TARE] to establish zero.

DISPLAY READS 0.00

5. Pause for motion delay.

DISPLAY READS Enter ~ Load ~ 0.00

6. Place a 100lb test weight on scale.

DISPLAY READS Enter ~ Load ~ 99.66

7. Enter

100.

DISPLAY READS 100

8. Press

[TARE] to establish span.

DISPLAY READS 100.00

9. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 100.00

10. Press [TARE] to accept calibration.

DISPLAY READS Enter ~ =Stor

11. Press [TARE] to save calibration.

DISPLAY READS Enter ~ =End

12. Press [TARE] to exit calibration.

DISPLAY READS 100.00

13. Remove the calibration weight.

DISPLAY READS 0.00

First Zero Calibration Method Example (375):

1. From the Weigh Mode key in 100 [SELECT].