79

2.5 V

8. Press

[SAMPLE/Enter] to increase the output to 50%.

0-20A ~ 50P
5 V

9. Press

[SAMPLE/Enter] to increase the output to 75%.

0-20A ~ 75P
7.5 V

10. Press

[SAMPLE/Enter] to increase the output to 100%.

0-20A ~ 100P
10 V

To test the 4-20mA output mode:



Voltmeter readings are based on the use of a 500 ohm precision resistor. Caution! Do not exceed
500 ohms.

1.

Enter the Setup Mode (see Setup Mode on page 15).
Chngs Poss!
P110.-- ~ F.S.= ~ 100.00

2.

Attach the voltmeter + (red) lead to pin 3 (0-10VDC) and the - (black) lead to pin 2
(ISOLATED GND) of the Analog Output connector.

3. Key

in 62004 [SELECT].

Test ~ 4-20A
Per P176

4. Press

[SAMPLE/Enter] to set the output to 0%.

4-20A ~ 0P
2 V

5. Press

[SAMPLE/Enter] to increase the output to 25%.

4-20A ~ 25P
4 V

6. Press

[SAMPLE/Enter] to increase the output to 50%.

4-20A ~ 50P
6 V

7. Press

[SAMPLE/Enter] to increase the output to 75%.

4-20A ~ 75P
8 V

8. Press

[SAMPLE/Enter] to increase the output to 100%.

4-20A ~ 100P
10 V