77

PARAMETER

NAME

DESCRIPTION

61117

61120

AiN1 ~ NrOff
-11035

AiN8 ~ NrOff
-14800

A/D non-ratio-metric offset compensation.

61121

Vre f~ NrOff
-12739

A/D reference voltage compensation.

61200

10oFF

Analog option 0 – 10V Zero offset value.

61201

10Gn

Analog option 0 – 10V Full scale gain value.

61202

0oFF

Analog option 0 – 20mA Zero offset value.

61203

0Gn

Analog option 0 – 20mA Full scale gain value.

61204

4oFF

Analog option 4 – 20mA Zero offset value.

61205

4Gn

Analog option 4 – 20mA Full scale gain value.

61206

Srln

Analog option board serial number.

62000

Dsply ~ Test
8.8.8.8.8.8.

Display test. Press [TARE] or [SAMPLE/Enter] to illuminate all segments.
Continue pressing [TARE] or [SAMPLE/Enter] to cycle through various
patterns.

62001

Spt 1 ~ Disbl

62002

Spt 2 ~ Disbl

62003

Spt 3 ~ Disbl

Allows setpoint status to be changed by pressing [TARE] or
[SAMPLE/Enter] while viewing this parameter. Requires that setup was
entered using the access code.

62004

Analg ~ 0- 10v

62005

Analg ~ 0-20A

62006

Analg ~ 4-20A

Allows the analog output to be changed by pressing [TARE] or
[SAMPLE/Enter]. Output will toggle through 0, 25, 50 and 100 percent
while viewing this parameter. Requires that setup was entered using the
access code (see Analog Board Diagnostic and Test Procedures on page
77).

64000

Send ~ Setup

Transmits all setup information out the communication port.

64100

LnCnt ~ 0

Received setup line count.

64101

ErCnt ~ 0

Received setup error count.

64102

1stEr ~ None!

Parameter of the first setup receive error.

65001

Deflt ~ All

Default All. Sets all parameters to factory default settings. Press [TARE]
or [SAMPLE/Enter] to initiate default.

65002

Deflt ~ -CAL

Same as above, except calibration is retained.

A

A

/

/

D

D

C

C

a

a

l

l

i

i

b

b

r

r

a

a

t

t

i

i

o

o

n

n

P

P

r

r

o

o

c

c

e

e

d

d

u

u

r

r

e

e

The Model 370/375 Analog-to-Digital Converter (A/D) is calibrated at the factory to ensure a
stable, linear response to the load cell signal. This calibration procedure calculates critical values
that are permanently stored in parameters P61110 - P61121. The A/D calibration should not be
confused with the standard weight calibration. It should never be necessary to recalibrate the
A/D. However, if the values stored at parameters P61110 - P61121 appear to be reset to
0.00000 and/or 1.00000, then A/D recalibration is necessary. Contact GSE Scale Systems or
your local authorized GSE distributor for more information on this procedure.

A

A

n

n

a

a

l

l

o

o

g

g

B

B

o

o

a

a

r

r

d

d

D

D

i

i

a

a

g

g

n

n

o

o

s

s

t

t

i

i

c

c

s

s

A

A

n

n

d

d

T

T

e

e

s

s

t

t

P

P

r

r

o

o

c

c

e

e

d

d

u

u

r

r

e

e

s

s



The following test procedures affect the analog output signal levels. Be sure to disconnect all
peripheral devices attached to the analog option card.



Test equipment needed: precision DC voltmeter, 500 ohm precision resistor. The 500 ohm
resistor must meet the following specifications: .01% tolerance and 5ppm temperature
coefficient.