You can enter the Calibration Mode after accessing the Setup Mode to view and/or change
parameter settings (see Setup Mode on page 15).

To access the Calibration Mode when viewing any setup parameter:

1. From the Setup Mode, press [ZERO].

DISPLAY READS Enter ~ =Cal!

2. Press

[TARE] or [SAMPLE/Enter].

DISPLAY READS First ~ Zero? ~ -0.26

F

AST

C

ALIBRATION

Fast Calibration allows calibration of the Model 370/375 scale system without accessing the
Setup Mode.

To access Fast Calibration from the Weigh Mode (370):

1. From the Weigh Mode, press [ZERO] + [SELECT].

DISPLAY READS Setup~Enter ~ Code!

2. Press

[ZERO] [PRINT] [UNITS] [TARE].

DISPLAY READS Fast ~ Cal!~First ~ Zero? ~ -0.26

To access Fast Calibration from the Weigh Mode (375):

1. From the Weigh Mode key in 100 [SELECT].

DISPLAY READS Setup~Enter ~ Code!

2. Key

in

54321 [SAMPLE/Enter].

DISPLAY READS Fast ~ Cal!~First ~ Zero? ~ -0.26

Fast Calibration can also be accessed if the following data stream is received via the comm port:

100%s54321%e

P

ERFORMING

C

ALIBRATION

Calibration always begins by establishing a zero (no-load) reference. A complete calibration also
requires establishing a span (test load) reference. This section details various methods for
obtaining zero and span references.

Establishing Zero

The Model 370/375 provides five methods for obtaining a zero (no load) calibration reference,
First Zero, Last Zero, False Zero, Only Zero, and Cal Reset.



Press [ZERO] or [CLR] during calibration to back up one step in the procedure.