7. Enter

2500.

DISPLAY READS 2500

8. Press

[SAMPLE/Enter] to establish span.

DISPLAY READS 2500.

9. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 2500.

10. Press [SAMPLE/Enter] to accept calibration.

DISPLAY READS Enter ~ =Stor

11. Press [SAMPLE/Enter] to save calibration.

DISPLAY READS Enter ~ =End

12. Press [SAMPLE/Enter] to exit calibration.

DISPLAY READS 5055.

13. Remove the calibration weight.

DISPLAY READS 00.

O

NLY

Z

ERO

Only Zero is used to establish a new calibration zero without affecting the span. This is useful for
correcting changes to the scale's dead load, for example adding safety rails to a truck scale
platform.

Only Zero Calibration Example (370):

1. From the Weigh Mode, press [ZERO] + [SELECT].

DISPLAY READS Setup

2. Press

[ZERO] [PRINT] [UNITS] [TARE].

DISPLAY READS Fast ~ Cal~First ~ Zero? ~2640.

3. Press

[SELECT] [SELECT] [SELECT].

DISPLAY READS Only ~ Zero? ~ 2640.

4. Remove any load on the scale.

DISPLAY READS Only ~ Zero? ~ 2620.

5. Press

[TARE] to establish zero.

DISPLAY READS 00.

6. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 00.

7. Press

[TARE] to accept calibration.

DISPLAY READS Enter ~ =Stor

8. Press

[TARE] to save calibration.

DISPLAY READS Enter ~ =End

9. Press

[TARE] to exit calibration.

DISPLAY READS 00.

Only Zero Calibration Example (375):

1. From the Weigh Mode, press 100 [SELECT].

DISPLAY READS Setup

2. Press

54321 [SAMPLE/Enter].

DISPLAY READS Fast ~ Cal~First ~ Zero? ~2640.

3. Press

[SELECT] [SELECT] [SELECT].