Entering calibration values – Holtgreven GSE-375 User Manual

Page 78

Advertising
background image

72

process. Each analog option board can be identified by its serial number, which is entered in the
unit during the calibration process.

Table 4-1: Analog Output Calibration Parameters

Parameter
Setting

Displayed
Name

Description

Type/choices (*=Default)


P61200


10oFF

Value required to precisely output 0V in 0 –
10V output mode (i.e. offset).

Numeric Entry:
0 to 15,000
2,923*


P61201


10Gn

Value required to precisely output 10V in 0 –
10V output mode (i.e. gain).

Numeric Entry:
50,000 to 65,535
61,027*


P61202

0oFF

Value required to precisely output 0 mA in 0 –
20 mA output mode (i.e. offset).

Not adjustable:
0*


P61203


0Gn

Value required to precisely output 20 mA in 0 –
20mA output mode (i.e. gain).

Numeric Entry:
50,000 to 65,535
54,555*


P61204


4oFF

Value required to precisely output 0 mA in 4 –
20 mA output mode (i.e. offset).

Numeric Entry:
0 to 15,000
10,910*


P61205


4Gn

Value required to precisely output 20 mA in 4 –
20mA output mode (i.e. gain).

Numeric Entry:
50,000 to 65,535
54,555*


P61206

Srln

Analog option board serial number.

Numeric Entry:
0* - 4,294,967,295

Entering Calibration Values

Analog calibration values can be entered into the Model 370 and Model 375 by keying in the data
in the left-hand column, beginning at the line starting with "61200…", replacing the "%s" character
pairs with the [SELECT] key and "%e" with the [TARE] (370) or [SAMPLE] (375) key. The line
with the "%c" is not adjustable and cannot be entered.

You can adjust the included factory recommended offset and gain values to precisely configure
the system being connected to these outputs. Once the initial factory values have been entered,
pressing the [TARE] (370) or [SAMPLE] (375) key will cause the count value to increase one
count, and pressing the [PRINT] + [UNITS] keys will decrement the count value by one count. All
changes made are updated "live" to aid in calibrating a specific device to the analog option board
output. Holding down the key(s) will repeat the increment/decrement action.

Printout Example

100%s23640%e

Access Setup Modes, Allowing Changes


60100%s%e

P60100. c2004-GSE-

60101%s%e

P60101. 0Model 370 370p27132

60102%s%e

P60102. 09-02-2005


60200%s%e

P60200. BrdSn 420679

60201%s%e

P60201. AuditTrail Euro 00001

60202%s%e

P60202. InsSn102245

60203%s%e

P60203. AuditTrail Cal. 00025

60204%s%e

P60204. AuditTrail Setup 00050


61200%s2923%e

P61200. 10off 2923

61201%s61027%e

P61201. 10 Gn 61027

61202%s0%c

P61202. 0 off 0

61203%s54555%e

P61203. 0 Gn 54555

61204%s10910%e

P61204. 4 off 10910

61205%s54555%e

P61205. 4 Gn 54555

61206%s123456%e

P61206. Srl n 123456

%z Exit Setup Mode