Holtgreven GSE-375 User Manual

Page 26

Advertising
background image

20

3. Press

[CLR] to exit without saving changes.

DISPLAY READS Enter ~ =Undo

4. Press

[SAMPLE/Enter] to undo changes.

DISPLAY READS Enter ~ =End

5. Press

[SAMPLE/Enter] to complete exit.

DISPLAY READS 0.00

F

F

a

a

c

c

t

t

o

o

r

r

y

y

D

D

e

e

f

f

a

a

u

u

l

l

t

t

Parameter 65001 and 65002 are available to return the Model 370/375 to factory settings.
Parameter 65001 will reset parameters including the calibration, while parameter 65002 resets all
parameters except the calibration.

After a factory default is performed and saved the Model 370/375 will be in the Quantity mode.
The display will show ‘do APS’ which means an average piece weight needs to be established.
Either perform a sample (refer to page 50) or press [SELECT] to chose another mode.

Return to factory default (370):

1. Access the setup mode. See page 15.
2. Key

in

65001 or 65002 [SELECT]

DISPLAY READS P65002~default-Cal

3. Press

[TARE] once.

DISPLAY READS Are U sure?~default=Enter

4. Press

[TARE] once.

DISPLAY READS Setup done~P65002~default-Cal

5. Press

[ZERO] [ZERO] [TARE] [TARE]

DISPLAY READS do APS

Return to factory default (375):

1. Access the setup mode. See page 15.
2. Key

in

65001 or 65002 [SELECT]

DISPLAY READS P65002~default-Cal

3. Press

[SAMPLE/Enter] once.

DISPLAY READS Are U sure?~default=Enter

4. Press

[SAMPLE/Enter] once.

DISPLAY READS Setup done~P65002~default-Cal

5. Press

[ZERO] [CLR] [SAMPLE/Enter] [SAMPLE/Enter]

DISPLAY READS do APS

L

L

i

i

s

s

t

t

o

o

f

f

P

P

a

a

r

r

a

a

m

m

e

e

t

t

e

e

r

r

s

s

The Model 370 and Model 375 have several parameters that can be configured to your specific
application. Table 3-1 is a complete list of the available parameters. Also refer to page 23

for

explanations of each parameter. The parameters shaded in gray do not apply to the Model
370/375.