Holtgreven GSE-375 User Manual

Page 70

Advertising
background image

64

7. Press

[TARE] to use last zero.

DISPLAY READS Enter ~ Load? ~ 9970

8. Enter

10000.

DISPLAY READS 10000

9. Press

[TARE] to establish span.

DISPLAY READS 10000.

10. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 10000.

11. Press [TARE] to accept calibration.

DISPLAY READS Enter ~ =Stor

12. Press [TARE] to save calibration.

DISPLAY READS Enter ~ =End

13. Press [TARE] to exit calibration.

DISPLAY READS 10000.

14. Remove the calibration weight.

DISPLAY READS 0.

Last Zero Calibration With Weight Already Applied Example (375):

1. Remove any load on the scale.

DISPLAY READS 10.

2. Press

[ZERO] to zero the scale.

DISPLAY READS 00.

3. Apply a 10000 lb test weight to verify calibration.

DISPLAY READS 9970.

4. Press

100 [SELECT].

DISPLAY READS Setup

5. Press

54321 [SAMPLE/Enter].

DISPLAY READS Fast ~ Cal~First ~ Zero? ~ 9930.

6. Press

[SELECT].

DISPLAY READS Last ~ Zero? ~9930.

7. Press

[SAMPLE/Enter] to use last zero.

DISPLAY READS Enter ~ Load? ~ 9970.

8. Enter

10000.

DISPLAY READS 10000

9. Press

[SAMPLE/Enter] to establish span.

DISPLAY READS 10000.

10. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 10000.

11. Press [SAMPLE/Enter] to accept calibration.

DISPLAY READS Enter ~ =Stor

12. Press [SAMPLE/Enter] to save calibration.

DISPLAY READS Enter ~ =End

13. Press [SAMPLE/Enter] to exit calibration.

DISPLAY READS 10000.

14. Remove the calibration weight.

DISPLAY READS 00.