Eset, Alibration – Holtgreven GSE-375 User Manual

Page 73

Advertising
background image

67

DISPLAY READS Only ~ Zero? ~ 2640.

4. Remove any load on the scale.

DISPLAY READS Only ~ Zero? ~ 2620.

5. Press

[SAMPLE/Enter] to establish zero.

DISPLAY READS 00.

6. Pause for motion delay.

DISPLAY READS Cal ~ Good? ~ 00.

7. Press

[SAMPLE/Enter] to accept calibration.

DISPLAY READS Enter ~ =Stor

8. Press

[SAMPLE/Enter] to save calibration.

DISPLAY READS Enter ~ =End

9. Press

[SAMPLE/Enter] to exit calibration.

DISPLAY READS 00.

R

ESET

C

ALIBRATION

Cal Reset may be necessary when an over-load or under-load condition exists, preventing the
completion of the calibration process. Calibration Reset adjusts the zero and gain factors of the
A/D amplifier to factory default values for maximum sensitivity.

After performing a calibration reset, a complete recalibration is required. The effects of a
calibration reset do not take effect until the Model 370/375 recalibrated and calibration information
has been saved.



If Code 02 (under-load) or Code 03 (over-load) is displayed during calibration, press [CLR] to
perform a calibration reset.

Reset Calibration Gain Factors Example (370):

1. From the Weigh Mode, press [ZERO] + [SELECT].

DISPLAY READS Setup

2. Press

[ZERO] [PRINT] [UNITS] [TARE].

DISPLAY READS Fast ~ Cal~First ~ Zero? ~2640.

3. Remove any load on the scale.

DISPLAY READS First ~ Zero? ~ xx.xx

4. Press

[TARE] to establish zero.

DISPLAY READS Enter ~ Load? ~ 0.00

5. Place a 100lb test weight on scale.

DISPLAY READS Code03

6. Press

[ZERO].

DISPLAY READS First ~ Zero? ~ -0.26

7. Remove any load on the scale.

DISPLAY READS First ~ Zero? ~ -0.42

8. Press

[TARE] to establish zero.

DISPLAY READS 0.00

9. Pause for motion delay.

DISPLAY READS Enter ~ Load ~ 0.00

10. Place a 100lb test weight on scale.