KEYENCE BL-600 Series User Manual

Page 66

Advertising
background image

Chapter 4 Setup Software

56

4

5. Sending/reading result

The following messages appear to indicate whether the sending/receiving
operation was successful or whether it failed.

If the operation failed, check the following points:

Check that the BL-600 Series is in the “setting data send/receive waiting
status” during which the 1st, 3rd, and 5th STABILITY LEDs from the top
flash simultaneously.

Check that [[Com port]] is set as shown in (3).

Check that the special power supply is turned on.

Check that the RS-232C cable pin assignment of the BL-600 is the same as
that of the host computer.

➮ See pages 21, 28 and 35.

6. Canceling the “setting data send/receive waiting status”.

Press the TEST switch once. The “setting data send/receive waiting status” is
canceled.

Sending succeeded

Sending failed

Reading succeeded

Reading failed

LASER ON

OK/NG

TIMING

TEST

Press the TEST switch once.