Bp channel 4: dynamic waveforms, Bp channel 1: respiration artifact – Fluke Biomedical MPS450 User Manual

Page 102

Advertising
background image

MPS450

Operators Manual

4-20

BP Channel 4: Dynamic Waveforms

Table 4-34 lists the numeric control codes for the BP channel 4: dynamic waveforms

functions.

Table 4-34. Numeric Codes for BP Channel 4: Dynamic Waveforms Actions

Numeric

Control Code

Action

Remote Control

Entry

080

Run BP3 arterial dynamic wave 120/80.

P3ART

081

Run BP3 radial art dynamic wave 120/80.

P3RART

082

Run BP3 left vent dynamic wave 120/0.

P3LV

083

Run BP3 right atrium CVP dynamic wave 15/10.

P3CVP

084

Run BP3 pulmonary art dynamic wave 25/10.

P3PA

085

Run BP3 pulm art wedge dynamic wave 10/2.

P3W

086

Run BP3 left atrium dynamic wave 14/4.

P3LA

087

Run BP3 right vent dynamic wave 25/0.

P3RV

BP CHANNEL 4: DYNAMIC WAVEFORMS

378

Run BP4 right atrium CVP dynamic wave 15/10.

P4CVP

379

Run BP4 right vent dynamic wave 25/0.

P4RV

380

Run BP4 pulmonary art dynamic wave 25/10.

P4PA

381

Run BP4 pulm art wedge dynamic wave 10/2.

P4W

088

Start Swan-Ganz auto.

STSGAUTO

416

Start Swan-Ganz manual.

STSG

417

Insert (in Swan-Ganz manual).

INS

418

Inflate (in Swan-Ganz manual).

INF

419

Deflate (in Swan-Ganz manual).

DEF

420

Pull Back (in Swan-Ganz manual).

PLBK

BP Channel 1: Respiration Artifact

Table 4-35 lists the numeric control codes for the BP channel 1: respiration artifact

functions.

Table 4-35. Numeric Codes for BP Channel 1: Respiration Artifact Actions

Numeric

Control Code

Action

Remote Control

Entry

349

Set BP1 respiration artifact off.

P1AOFF

350

Set BP1 respiration artifact on.

P1AON