Other Functions

Temperature (Adjusting Body Temperature)

3

3-7

within

±

0.40

°

C: freezing (0

°

C / 32

°

F); hypothermic (24

°

C / 75.2

°

F); normal (37

°

C /

98.6

°

F); or pyretic (40

°

C / 104

°

F).

Action in the Menu-Control Mode

1. Press the top-menu key labeled TEMP to display the following LCD screen:

gje037.eps

2. To adjust the simulated temperature to the desired setting, select DOWN or UP.
3. The temperature setting identified on the LCD screen is active when displayed and

remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for a Celsius-

temperature setting according to Table 3-6, and select RUN:

Table 3-6. Numeric Codes for Temperature-Celsius Settings

Temperature-Celsius Setting

Numeric Code

0 DEG

189

24 DEG

190

37 DEG

191

40 DEG

192