Apnea simulation, Blood pressure functions, Blood-pressure sensitivity – Fluke Biomedical MPS450 User Manual

Page 98: Blood pressure zeroing, Bp channel 1: static-pressure levels

Advertising
background image

MPS450

Operators Manual

4-16

Apnea Simulation

Table 4-24 lists the numeric control codes for the apnea simulation functions.

Table 4-24. Numeric Codes for Apnea Simulation Actions

Numeric

Control Code

Action

Remote Control

Entry

152

Start apnea for 12 seconds.

A12

153

Start apnea for 22 seconds.

A22

154

Start apnea for 32 seconds.

A32

150

Turn on apnea continuously.

AON

151

Turn off apnea.

AOFF

Blood Pressure Functions

The Blood Pressure Functions consists of Blood-Pressure Sensitivity, Blood Pressure

Zeroing, BP Channel 1: Static-Pressure Levels, BP Channel 2: Static-Pressure Levels, BP

Channel 3: Static-Pressure Levels, BP Channel 4: Static-Pressure Levels, BP Channel 1:

Dynamic Waveforms, BP Channel 2: Dynamic Waveforms, BP Channel 3: Dynamic

Waveforms, BP Channel 4: Dynamic Waveforms, BP Channel 1: Respiration Artifact,

BP Channel 2: Respiration Artifact, BP Channel 3: Respiration Artifact, and BP

Channel 4: Respiration Artifact.

Blood-Pressure Sensitivity

Table 4-25 lists the numeric control codes for the blood-pressure sensitivity functions.

Table 4-25. Numeric Codes for Blood-Pressure Sensitivity Actions

Numeric

Control Code

Action

Remote Control

Entry

069

Set blood-pressure sensitivity to 5

μ

V/V/mmHg. BPSNS5

068

Set Set blood-pressure sensitivity to 40

μ

V/V/mmHg. BPSNS40

132

Store blood-pressure sensitivity at 5

μ

V/V/mmHg. (N/A)

133

Store blood-pressure sensitivity to 40

μ

V/V/mmHg. (N/A)

(N/A)

Store the blood-pressure sensitivity

BPSNSSTORE

Blood Pressure Zeroing

Table 4-26 lists the numeric control codes for the blood-pressure zeroing functions.

Table 4-26. Numeric Codes for Blood-Pressure Zeroing Actions

Numeric

Control Code

Action

Remote Control

Entry

007

Zero all four blood pressure channels.

ZALL

BP Channel 1: Static-Pressure Levels

Table 4-27 lists the numeric control codes for the BP channel 1: static-pressure levels

functions.