Performance waves, Performance wave amplitude – Fluke Biomedical MPS450 User Manual

Page 93

Advertising
background image

Remote Operations

Codes and Actions

4

4-11

Performance Waves

Table 4-15 lists the numeric control codes for the performance waves functions.

Table 4-15. Numeric Codes for Performance Waves Actions

Numeric

Control Code

Action

Remote Control

Entry

120

Run 2 Hz square performance wave.

SQU2

121

Run .125 Hz square performance wave.

SQU.125

Table 4-15. Numeric Codes for Performance Waves Actions (cont.)

Numeric

Control Code

Action

Remote Control

Entry

122

Run 2 Hz triangle performance wave.

TRI2

123

Run 2.5 Hz triangle performance wave.

TRI2.5

124

Run 30 BPM pulse performance wave.

PUL30

125

Run 60 BPM pulse performance wave.

PUL60

207

Run 0.5 Hz sine performance wave.

SIN0.5

208

Run 5 Hz sine performance wave.

SIN5

209

Run 10 Hz sine performance wave.

SIN10

210

Run 40 Hz sine performance wave.

SIN40

211

Run 50 Hz sine performance wave.

SIN50

212

Run 60 Hz sine performance wave.

SIN60

213

Run 100 Hz sine performance wave.

SIN100

Performance Wave Amplitude

Table 4-16 lists the numeric control codes for the performance wave amplitude functions.

Table 4-16. Numeric Codes for Performance Wave Amplitude Actions

Numeric

Control Code

Action

Remote Control

Entry

272

Set performance amplitude to 0.05 mV.

PFA0.05

273

Set performance amplitude to 0.10 mV.

PFA0.10

274

Set performance amplitude to 0.15 mV.

PFA0.15

275

Set performance amplitude to 0.20 mV.

PFA0.20

276

Set performance amplitude to 0.25 mV.

PFA0.25

277

Set performance amplitude to 0.30 mV.

PFA0.30

278

Set performance amplitude to 0.35 mV.

PFA0.35

279

Set performance amplitude to 0.40 mV.

PFA0.40

280

Set performance amplitude to 0.45 mV.

PFA0.45