Viewing current mps450 parameters – Fluke Biomedical MPS450 User Manual

Page 24

Advertising
background image

MPS450

Operators Manual

1-10

Table 1-3. MPS450 Button Description (cont.)

Number Key /

Key

Menu Label

Menu Functions

SETUP

Set respiration lead and baseline (impedance).
Set blood-pressure sensitivity.
Adjust viewing angle (contrast) for LCD screen.
Set beeper mode.

Switch between menu control and numeric control.

View current parameters.

Return to previous or top menu.

Select option displayed on LCD screen.

Viewing Current MPS450 Parameters

The current settings for adjustable parameters—such as heart rate and BP static-pressure

levels—are available for display at any time on a series of LCD screens that are accessed

by pressing the VIEW key.
(When the MPS450 is turned off, the parameters reset to defaults, with the exception of

settings that can be stored, i.e., respiration lead and baseline, BP sensitivity, view angle,

and beeper mode. Power-on default settings for functions are listed in the

“Specifications” section later in this chapter.)

Action in the Menu-Control Mode

1. Press the white key labeled VIEW to display the VIEW PAGE 1 screen—the ECG

page—which indicates current ECG settings:

gje003.eps

The types of parameters displayed on VIEW PAGE 1 depend on which group of

ECG waves is running. In this example (the ECG-NSR-wave group), the settings

displayed include NSR rate (90 BPM), patient type (ADULT), ST-segment elevation

(+.30 mV), NSR amplitude (2.00 mV), and ECG artifact (50 HZ).
Four other types of ECG-wave groups display different parameter types on VIEW

PAGE 1, and three other VIEW PAGE screens display parameters for respiration,

blood-pressure, and temperature/cardiac-output functions. Table 1-4 lists the

parameters available for viewing on the four pages: