Pacemaker waveform, Pacemaker amplitude, Pacemaker width – Fluke Biomedical MPS450 User Manual

Page 90

Advertising
background image

MPS450

Operators Manual

4-8

Pacemaker Waveform

Table 4-8 lists the numeric control codes for the pacemaker waveform functions.

Table 4-8. Numeric Codes for Pacemaker Waveform Actions

Numeric

Control Code

Action

Remote Control

Entry

383

Run atrial pacer wave.

ATR

110

Run asynchronous pacer wave.

ASN

111

Run demand frequent sinus pacer wave.

DFS

112

Run demand occasional sinus pacer wave.

DOS

113

Run AV sequential pacer wave.

AVS

114

Run noncapture pacer wave.

NCA

115

Run nonfunction pacer wave.

NFU

Pacemaker Amplitude

Table 4-9 lists the numeric control codes for the pacemaker amplitude functions.

Table 4-9. Numeric Codes for Pacemaker Amplitude Actions

Numeric

Control Code

Action

Remote Control

Entry

384

Set pacer amplitude to 1 mV.

PA1

385

Set pacer amplitude to 2 mV.

PA2

386

Set pacer amplitude to 5 mV.

PA5

242, 387

Set pacer amplitude to 10 mV.

PA10

Pacemaker Width

Table 4-10 lists the numeric control codes for the pacemaker width functions.

Table 4-10. Numeric Codes for Pacemaker Width Actions

Numeric

Control Code

Action

Remote Control

Entry

243

Set pacer width to 0.1 ms.

PW0.1

244

Set pacer width to 0.5 ms.

PW0.5

245

Set pacer width to 1.0 ms.

PW1.0

246

Set pacer width to 1.5 ms.

PW1.5

247

Set pacer width to 2.0 ms.

PW2.0