Respiration functions, Respiration lead, Respiration baseline (impedance) – Fluke Biomedical MPS450 User Manual

Page 96

Advertising
background image

MPS450

Operators Manual

4-14

Table 4-19. Numeric Codes for R-Wave Amplitude Actions (cont.)

Numeric

Control Code

Action

Remote Control

Entry

302

Set R-wave amplitude to 1.00 mV.

RWA1.00

303

Set R-wave amplitude to 1.50 mV.

RWA1.50

304

Set R-wave amplitude to 2 mV.

RWA2.00

305

Set R-wave amplitude to 2.50 mV.

RWA2.50

306

Set R-wave amplitude to 3.00 mV.

RWA3.00

307

Set R-wave amplitude to 3.50 mV.

RWA3.50

308

Set R-wave amplitude to 4.00 mV.

RWA4.00

309

Set R-wave amplitude to 4.50 mV.

RWA4.50

310

Set R-wave amplitude to 5.00 mV.

RWA5.00

311

Set R-wave amplitude to 5.50 mV.

RWA5.50

Respiration Functions

The Respiration Functions category consists of Respiration Lead, Respiration Baseline

(Impedance), Respiration Rate, Respiration Amplitude, and Apnea Simulation.

Respiration Lead

Table 4-20 lists the numeric control codes for the respiration lead functions.

Table 4-20. Numeric Codes for Respiration Lead Actions

Numeric

Control Code

Action

Remote Control

Entry

(N/A)

Store the respiration lead.

RLSTORE

389

Set respiration lead to LA.

RLLA

391

Set respiration lead to LL.

RLLL

390

Store LA respiration lead.

(N/A)

392

Store LL respiration lead.

(N/A)

Respiration Baseline (Impedance)

Table 4-21 lists the numeric control codes for the respiration baseline (impedance)

functions.

Table 4-21. Numeric Codes for Respiration Baseline (Impedance) Actions

Numeric

Control Code

Action

Remote Control

Entry

180

Set respiration base to 500 ohms.

RB500

181

Set respiration base to 1000 ohms.

RB1000

182

Set respiration base to 1500 ohms.

RB1500

183

Set respiration base to 2000 ohms.

RB2000