Fluke Biomedical MPS450 User Manual

Page 108

Advertising
background image

MPS450

Operators Manual

A-2

Table A-1. Problem vs. Solution (cont.)

Problem Probable

Cause

Solution

Respiration

Respiration lead not
selected correctly

Check monitor respiration-lead detection and
switch accordingly on the MPS450.

Leads not connected
properly

Verify that the color code is correct for U.S.
or International ECG leads.

No respiration signal

Performance waveform
currently output (which
turns off respiration)

Select normal physiological waveform.

Temperature

No temperature reading

Wrong series
temperature cable used

Check manufacturer’s series for thermistor
used. Choose the correct Fluke Biomedical
temperature cable for 700 or 400 Series.

Cardiac Output

Selected Injectate
temperature not set
correctly

Injectate temperature must match expected
temperature: 24

°

C or 0

°

C. Use optional

cardiac-output box for setting injectate
temperature.

Incorrect setting for
injectate volume on
cardiac output computer

Set to 10 cc.

Incorrect setting for
catheter size on cardiac
output computer

Set to 7f.

Expected values not correct, or
no values

Correct constant not set
on monitor

Set correct constant to .542 or .595. See
chapter on Cardiac Output.

Injectate temperature not
displayed on monitor

Turn on injectate-temperature pot. (Zero the
box until reading displayed is 0

°

C or 24

°

C.)