Adjusting pacemaker–spike amplitude, Adjusting pacemaker–spike width – Fluke Biomedical MPS450 User Manual

Page 41

Advertising
background image

Cardiac Functions

ECG Functions

2

2-9

Alternatively, in the Numeric-Control Mode, press the number keys for a pacemaker-type wave
according to Table 2-6, and select RUN:

Table 2-6. Numeric Code for Pacemaker-Waveform Settings

Pacemaker-Waveform Setting

Numeric Code

ATRIAL 80 BPM

383

ASYNC 75 BPM

110

DEMAND FREQ SINUS

111

DEMAND OCC SINUS

112

AV SEQUENTIAL

113

NON-CAPTURE 114

NON-FUNCTION 115

Adjusting Pacemaker–Spike Amplitude

The MPS450 offers a selection of four preprogrammed settings (mV) for pacemaker-spike
amplitude: 1, 2, 5, or 10.

Action in the Menu-Control Mode

1. On the LCD screen for the selected pacer-type wave, select AMPL repeatedly to

scroll to the desired spike-amplitude setting.

2. The pacemaker amplitude identified on the LCD screen is active when displayed and

remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for pacemaker-

spike amplitude according to Table 2-7, and select RUN:

Table 2-7. Numeric Codes for Pacemaker-Amplitude Settings

Pacemaker-Amplitude Setting

Numeric Code

1.0 mV

384

2.0 mV

385

5.0 mV

386

10.0 mV

387

Adjusting Pacemaker–Spike Width

The MPS450 simulates five preconfigured settings (ms) for pacemaker-spike width: 0.1,

0.5, 1.0, 1.5, or 2.0.

Action in the Menu-Control Mode

1. On the LCD screen for the selected pacer-type wave, select WIDTH repeatedly to

scroll to the desired spike-width setting.

2. The pacemaker width identified on the LCD screen is active when displayed and

remains active until the settings are changed.

Alternatively, in the Numeric-Control Mode, press the number keys for the pacemaker-

width setting according to Table 2-8, and select RUN: