Ecg-performance-testing, Respiration, Blood pressure – Fluke Biomedical MPS450 User Manual

Page 30

Advertising
background image

MPS450

Operators Manual

1-16

Ventricular Arrhythmia ......................................... PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy;

trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); ventricular
tachycardia; ventricular fibrillation (coarse or fine); and asystole

Conduction Defect ................................................ First-, second-, or third-degree heart block; and right- or left-bundle-

branch block

Power-On Default.................................................. Atrial fibrillation (coarse); PAC; PVCs 6/minute; first-degree heart block

ECG-Performance-Testing

Amplitude .............................................................. 0.05 to 0.5 mV (0.05 mV steps)

0.5 to 5.5 mV (0.5 mV steps)

Pulse Wave ............................................................ 30, 60 BPM, with 60 ms pulse width
Square Wave ......................................................... 2.0, 0.125 Hz
Triangle Wave........................................................ 2.0, 2.5 Hz
Sine Wave .............................................................. 0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave-Detection Waveform ............................... Haver-Triangle
R-wave Rate........................................................... 30, 60, 80, 120, 200, and 250 BPM
R-wave Width ........................................................ 20 to 200 ms (10 ms steps)

Additional Steps: 8, 10, and 12 ms

Rate Accuracy .......................................................

±

1 %

Amplitude Accuracy .............................................

±

2 %, Lead II (Exception:

±

5 % for R waves

20 ms)

Power-On Default ................................................. 1.0 mV, square wave 2.0 Hz, R-wave rate 60 BPM, R-wave width

10 ms

Respiration

Rate ........................................................................ 0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Impedance Variations (

Δ

Ω

) ................................. 0.2, 0.5, 1, or 3

Ω

Accuracy Delta ......................................................

±

10 %

Baseline ................................................................. 500, 1000, 1500, 2000

Ω

, Leads I, II, III

Accuracy Baseline ................................................

±

5 %

Respiration Lead................................................... LA or LL
Apnea Selection .................................................... 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON =

respiration OFF)

Power-On Default.................................................. 20 BrPM, delta 1.0

Ω

, 1000-

Ω

baseline, left-arm lead (LA), 12-second

apnea

Blood Pressure

Input/output Impedance ....................................... 300

Ω

±

10 %

Exciter Input Range .............................................. 2.0 to 16.0 V rms
Exciter-Input Frequency Range........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5 or 40

μ

V/V/mmHg

Pressure Accuracy ...............................................

±

(2 % of setting + 2 mmHg)

Static Levels, p1.................................................... -10, 0, 80, 160, 240, 320, 400 mmHg
Static Levels, p2.................................................... -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels, p3.................................................... -5, 0, 20, 40, 60, 80, 100 mmHg
Static Levels, p4.................................................... -5, 0, 20, 40, 60, 80, 100 mmHg
Dynamic Waveforms, p1 ...................................... Arterial: 120/80

Radial artery: 120/80
Left ventricle: 120/00
Right ventricle: 25/00

Dynamic Waveforms, p2 ...................................... Arterial: 120/80

Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4

Dynamic Waveforms, p3 ...................................... Arterial: 120/80

Radial artery: 120/80
Left ventricle: 120/00

Advertising