Running a dynamic waveform – Fluke Biomedical MPS450 User Manual

Page 63

Advertising
background image

Cardiac Functions

Blood Pressure Function

2

2-31

Running a Dynamic Waveform

The MPS450 simulates dynamic waves to represent the pressures in various places in the

heart or blood vessels. Waves are specified for a normal sinus rhythm at 80 BPM, with

systolic pressure over diastolic pressure, e.g., 120/80, etc. Available dynamic waves (not

all available on all channels) are listed Table 2-24. (For information about the Swan-Ganz

simulation—a serial simulation available only on BP4, in both manual and automatic

modes—see the “Simulating the Swan-Ganz Procedure” section in this chapter.

Table 2-24. Dynamic Waveform Availability by BP Channel

DYNAMIC

WAVE BP1 BP2 BP3 BP4

ARTERIAL 120/80

RADIAL
ARTERY

120/80

LEFT
VENTRICLE

120/00

LEFT ATRIUM 14/4

RIGHT
ATRIUM (CVP)

15/10

RIGHT
VENTRICLE

25/00

PULMONARY
ARTERY

25/10

PULMONARY
WEDGE

10/2

IN SWANZ-GA

NZ

PROCEDURE

Action in the Menu-Control Mode

1. Press the top-menu key labeled BP.
2. Select BP1, BP2, BP3, or BP4 to display the screen for a specific channel.
3. Select DYNA to display the DYNAMIC WAVE screen for the selected channel (in

this example, BP3):

gje030.eps

4. Scroll to the desired dynamic waveform, for example, PULM WEDGE 10/2.
5. Select RUN. On the LCD screen, RUN flashes to indicate the identified waveform is

running. The simulation runs continuously (repeats) until another selection is made.

6. Press the ESC key to return to the screen for the selected BP channel. The screen

identifies the selected waveform.

7. Press the ESC key again to return to the top menu BLOOD PRESSURE. The LCD

screen indicates the waveform running on the selected channel:

Advertising