Pacemaker waveforms – Fluke Biomedical MPS450 User Manual

Page 40

Advertising
background image

MPS450

Operators Manual

2-8

Table 2-5. Numeric Code for ECG Artifact Settings

ECG-Artifact Setting

Numeric Code

ECG ARTIFACT OFF

104

60 HZ

105

50 HZ

106

MUSCLE 107

WANDER 108

RESPIRATION 109

Pacemaker Waveforms

When the heart beats erratically, or not at all, it may need to be stimulated by artificial

means, that is, by either an internal (permanent) or an external (temporary) artificial

pacemaker. Pacemakers may operate at a fixed preset rate, or on demand.
The MPS450 sends waveforms to simulate several artificial-pacemaker conditions:

an atrial pacemaker wave at 80 BPM, with a pacer pulse at the start of each P

wave;

an asynchronous pacemaker wave with continuous ventricular-paced beats (75

BPM) and no P waves;

a “demand” pacemaker wave with frequent sinus beats (forty normal beats

followed by twenty ventricular-paced beats, repeated);

a “demand” pacemaker wave with occasional sinus beats (twenty normal beats

followed by forty ventricular-paced beats, repeated);

an AV-sequential-pacemaker wave with continuous paced beats, each with an

atrial pulse and a P wave followed by a ventricular-paced pulse and QRS

response;

ventricular-paced beats, where one out of every ten beats has no heart response

(noncapture); or

continuous pacer pulses at 75 BPM with no heart response (nonfunction).

Action in the Menu-Control Mode

1. Press the top-menu key labeled PACE to display the following LCD screen:

gje014.eps

2. Scroll to the desired pacemaker-wave type: ATRIAL 80 BPM; ASYNC 75 BPM;

DEMAND FREQ SINUS; DEMAND OCC SINUS; AV SEQUENTIAL; NON-

CAPTURE; or NON-FUNCTION.

3. The pacemaker wave identified on the LCD screen is active when displayed and

remains active until another wave is selected.

Advertising