Zeroing bp channels – Fluke Biomedical MPS450 User Manual

Page 60

Advertising
background image

MPS450

Operators Manual

2-28

gje025.eps

2. Select BP SENSE to display the BP SENSITIVITY screen. If the setting has not

been changed, the LCD screen displays the default BP-sensitivity setting (5

μ

V/V/mmHg):

gje026.eps

3. To change the BP-sensitivity setting, select 40. The setting identified on the LCD

screen is active when displayed and remains active until the setting is changed. (This

setting does not persist beyond the current session unless it is stored specifically.)

4. To store the displayed value as the power-on default, select STORE. On the LCD

screen Storing blinks on momentarily to indicate the value is being saved.

5. Press the ESC key to return to the top menu SETUP.
Alternatively, in the Numeric-Control Mode, press the number keys for a BP-sensitivity

setting according to Table 2-22, and select RUN:

Table 2-22. Numeric Codes for BP-Sensitivity Settings

BP-Sensitivity Setting

Numeric Code

40 uV/V

068

40 uV/V STORE

133

5 uV/V

069

5 uV/V STORE

132

Zeroing BP Channels

Before testing or simulation, each of the four BP channels should be reset to zero mmHg.

Action in the Menu-Control Mode

1. To display the active pressure setting for each of the four channels, press the top-

menu key labeled BP. (In this example, the active setting for BP channel 3 has been

set to 40 mmHg.)