Cardiac Functions

Blood Pressure Function

2

2-29

gje027.eps

2. To display the screen for BP channel 3, select BP3:

gje028.eps

3. To reset BP channel 3 to zero, select ZERO.
4. Press the ESC key to return to the top menu BLOOD PRESSURE.
5. For each channel, select its screen from the top menu, and then select ZERO, until

the top menu BLOOD PRESSURE displays 0 as the active setting for all four

channels.

Alternatively, in the Numeric-Control Mode, to zero all four BP channels simultaneously,

press the number keys 007, and select RUN.

Setting Static-Pressure Levels

The MPS450 offers seven preconfigured static-pressure-level settings (mmHg) for each

BP channel, with each channel controlled independently:

Set BP channel 1 to -10, 0, 80, 160, 240, 320, or 400.

Set BP channel 2 to -10, 0, 50, 100, 150, 200, or 240.

Set each of BP channels 3 and 4 to -5, 0, 20, 40, 60, 80, or 100.

Action in the Menu-Control Mode

1. Press the top-menu key labeled BP.
2. Select BP1, BP2, BP3, or BP4 to display the channel’s screen.
3. Select STAT to display the STATIC screen for the selected channel (in this example,

BP channel 1 with an active setting of 80 mmHg):

gje029.eps

4. On the LCD screen, RUN flashes, because the displayed setting is active.