General remote commands, Error messages, Codes and actions – Fluke Biomedical MPS450 User Manual

Page 86

Advertising
background image

MPS450

Operators Manual

4-4

The MPS450 executes a command upon a carriage return and/or a line feed.

Alphabetic characters in commands are case-insensitive.

When entering commands, the BACKSPACE key operates normally (that is, deletes

the previously recorded character).

When entering commands, the ESCAPE key discards the entire command.

When a command is completed, a message is sent back (usually “OK”).

Using Remote Entry to Operate In Numeric-Control Mode

In addition to using remote commands to set up wave parameters, one specific remote-

entry command can be used to operate the MPS450 in the numeric-control mode from a

peripheral device:

NUMENT=num

(where

num

is a numeric-control code)

For example, entering the remote command

NUMENT=017

runs a simulation of atrial

tachycardia—just as if the MPS450 was being operated in the numeric-control mode, the

number keys 017 were pressed on the MPS450 keypad, and RUN was selected.
A list of numeric-control codes—cross-referenced both to actions and to remote-entry

commands—is available in the “Codes and Actions” section later in this chapter.

General Remote Commands

The general remote commands are listed in Table 4-2:

Table 4-2. General Remote Commands

Remote Command

Action

IDENT

Identifies the instrument. Returns “MPS450; version; options,” where
version is the version of the firmware program and options equal “C” for
Cardiac Output, and/or “F” for Fetal / Maternal.

VER

Returns the firmware version.

NUMENT=

num

Executes a numerical entry function, where num equals the 3-digit number
to execute.

Error Messages

Invalid commands return error messages, including the following:

“ERR=00, NO COMMANDS ALLOWED AT THIS TIME”

“ERR=001, UNKNOWN COMMAND”

“ERR=02, ILLEGAL COMMAND”

“ERR=03, ILLEGAL PARAMETER”

“ERR=04, DATA CORRUPTED”

“ERR=05, UNKNOWN ERROR”

“ERR=06, OPTION NOT INSTALLED”

“ERR=20, INVALID NUMERIC ENTRY”

“ERR=21, ILLEGAL SWAN-GANZ COMMAND”

Codes and Actions

This section lists the code numbers for controlling MPS450 actions, including both

numeric-mode and remote-control-entry codes. (See the Table of Contents for this

chapter to locate a specific action.)

Advertising