Bp channel 2: respiration artifact, Bp channel 3: respiration artifact, Bp channel 4: respiration artifact – Fluke Biomedical MPS450 User Manual

Page 103: Other functions, Temperature

Advertising
background image

Remote Operations

Codes and Actions

4

4-21

BP Channel 2: Respiration Artifact

Table 4-36 lists the numeric control codes for the BP channel 2: respiration artifact

functions.

Table 4-36. Numeric Codes for BP Channel 2: Respiration Artifact Actions

Numeric

Control Code

Action

Remote Control

Entry

358

Set BP2 respiration artifact off.

P2AOFF

359

Set BP2 respiration artifact on.

P2AON

BP Channel 3: Respiration Artifact

Table 4-37 lists the numeric control codes for the BP channel 3: respiration artifact

functions.

Table 4-37. Numeric Codes for BP Channel 3: Respiration Artifact Actions

Numeric

Control Code

Action

Remote Control

Entry

367

Set BP3 respiration artifact off.

P3AOFF

368

Set BP3 respiration artifact on.

P3AON

BP Channel 4: Respiration Artifact

Table 4-38 lists the numeric control codes for the BP channel 4: respiration artifact

functions.

Table 4-38. Numeric Codes for BP Channel 4: Respiration Artifact Actions

Numeric

Control Code

Action

Remote Control

Entry

376

Set BP4 respiration artifact off.

P4AOFF

377

Set BP4 respiration artifact on.

P4AON

Other Functions

The Other Functions category consists of Temperature, Cardiac-Output Wave/Injectate,

FHR Rate (Fixed), Intrauterine-Pressure (-) Wave, Intrauterine-Pressure (IUP) Period,

View Angle, and Beeper.

Temperature

Table 4-39 lists the numeric control codes for the temperature functions.

Table 4-39. Numeric Codes for Temperature Actions

Numeric

Control Code

Action

Remote Control

Entry

189

Set temperature to 0 degrees C.

T0

190

Set temperature to 24 degrees C.

T24

191

Set temperature to 37 degrees C.

T37

192

Set temperature to 40 degrees C.

T40