Bp channel 2: static-pressure levels, Bp channel 3: static-pressure levels – Fluke Biomedical MPS450 User Manual

Page 99

Advertising
background image

Remote Operations

Codes and Actions

4

4-17

Table 4-27. Numeric Codes for BP Channel 1: Static-Pressure Levels Actions

Numeric

Control Code

Action

Remote Control

Entry

342

Set BP1 to -10 mmHg static.

P1S-10

343

Set BP1 to zero.

P1S0

344

Set BP1 to 80 mmHg static.

P1S80

345

Set BP1 to 160 mmHg static.

P1S160

346

Set BP1 to 240 mmHg static.

P1S240

347

Set BP1 to 320 mmHg static.

P1S320

348

Set BP1 to 400 mmHg static.

P1S400

BP Channel 2: Static-Pressure Levels

Table 4-28 lists the numeric control codes for the BP channel 2: static-pressure levels

functions.

Table 4-28. Numeric Codes for BP Channel 2: Static-Pressure Levels Actions

Numeric

Control Code

Action

Remote Control

Entry

351

Set BP2 to -10 mmHg static.

P2S-10

352

Set BP2 to zero.

P2S0

353

Set BP2 to 50 mmHg static.

P2S50

354

Set BP2 to 100 mmHg static.

P2S100

355

Set BP2 to 150 mmHg static.

P2S150

356

Set BP2 to 200 mmHg static.

P2S200

357

Set BP2 to 240 mmHg static.

P2S240

BP Channel 3: Static-Pressure Levels

Table 4-29 lists the numeric control codes for the BP channel 3: static-pressure levels

functions.

Table 4-29. Numeric Codes for BP Channel 3: Static-Pressure Levels Actions

Numeric

Control Code

Action

Remote Control

Entry

360

Set BP3 to -5 mmHg static.

P3S-5

361

Set BP3 to zero.

P3S0

362

Set BP3 to 20 mmHg static.

P3S20

363

Set BP3 to 40 mmHg static.

P3S40