Adult and pediatric ecg – Fluke Biomedical MPS450 User Manual

Page 37

Advertising
background image

Cardiac Functions

ECG Functions

2

2-5

Alternatively, in the Numeric-Control Mode, press the number keys for an ECG-

amplitude setting according to Table 2-2, and select RUN:

Table 2-2. Numeric Codes for ECG Amplitude Settings

Amplitude setting

Numeric code

0.05 mV

252

0.10 mV

253

0.15 mV

254

0.20 mV

255

0.25 mV

256

0.30 mV

257

0.35 mV

258

0.40 mV

259

0.45 mV

260

0.50 mV

261

1.00 mV

262

1.50 mV

263

2.00 mV

264

2.50 mV

265

3.00 mV

266

3.50 mV

267

4.00 mV

268

4.50 mV

269

5.00 mV

270

5.50 mV

271

Adult and Pediatric ECG

The MPS450 simulates an adult ECG waveform with QRS duration of 80 ms, or a

pediatric waveform with QRS duration of 40 ms.

Action

1. Press the top-menu key labeled NSR. The LCD screen identifies the current patient

type either as adult (ADULT) or as pediatric (PEDS).

2. Select > to move to the LCD screen PT TYPE.
3. Select PT TYPE to toggle to the desired patient type.
4. The patient-type setting identified on the LCD screen is active when displayed and

remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for a patient-type

setting according to Table 2-3, and select RUN: