Fluke Biomedical MPS450 User Manual

Page 18

Advertising
background image

MPS450

Operators Manual

1-4

Table 1-1. MPS450 Functions

Function Category

Function

Description

Normal sinus rhythm

ECG rate and amplitude

Adult / pediatric QRS

ST-segment elevation

ECG-artifact simulation

Pacemaker waveforms

ECG Functions

Pacer amplitude and
width

The MPS450 provides complete 3-, 5-, and
12-lead ECG simulation that includes seven
artificial pacemaker conditions. Normal sinus
rhythm is output over a range of heart rates
and voltage amplitudes.
ST-segment elevation is adjustable.

Supraventricular
arrhythmia

Premature arrhythmia

Missed beat

Ventricular arrhythmia

Arrhythmia Functions

Conduction defect

The MPS450 simulates 36 types of
arrhythmias, such as PVCs, tachycardia,
fibrillation, flutter, and asystole. Simulated
conduction defects include first-, second-, and
third-degree heart block; and left- and right-
bundle-branch block.

Square/pulse/triangle/sine

R waveforms

Wave amplitude

ECG-Performance Testing

R-wave rate and width

The MPS450 generates square, pulse,
triangle, sine, and R waveforms for
performance testing. Wave amplitude is
adjustable, as well as R-wave rate and width.

Respiration lead

Baseline (impedance)

Respiration rate

Respiration amplitude

Respiration

Apnea simulation

Calibrated respiration rates are generated
from 15 to 120 BrPM (breaths per minute),
including four respiration-impedance
selections, with two different lead selections
(LA or LL). The output-impedance level is
adjustable to 500, 1000, 1500, or 2000 ohms.
The MPS450 generates apnea pauses (0
BrPM) of 12, 22, and 32 seconds, as well as a
continuous-apnea condition.

BP sensitivity

BP zeroing

Static-pressure levels

Dynamic BP waveforms

BP respiration artifact

Blood Pressure

Swan-Ganz simulation

The MPS450 simulates static and dynamic
invasive pressures, providing complete blood-
pressure simulation. The MPS450 also
provides calibrated static pressures and
dynamic waveforms to simulate signals such
as pulmonary-artery, left- and right-ventricle
and Swan Ganz (RA-RV-PA-PAW) pressures.

Temperature Temperature

settings

The MPS450 provides four preset
temperature simulations: 0

°

C, 24

°

C, 37

°

C,

and 40

°

C. All temperature simulations are

compatible with Yellow Springs, Inc. (YSI)
Series 400 and 700 thermistors.