Fluke Biomedical MPS450 User Manual

Page 79

Advertising
background image

Other Functions

Respiration

3

3-5

2. To adjust the respiration rate, select DOWN or UP. The rate identified on the LCD

screen is active when displayed and remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-rate

setting according to Table 3-3, and select RUN:

Table 3-3. Numeric Codes for Respiration-Rate Settings

Respiration-Rate Setting

Numeric Code

0 BrPM

156

15 BrPM

157

20 BrPM

158

30 BrPM

159

40 BrPM

160

60 BrPM

161

80 BrPM

162

100 BrPM

163

120 BrPM

164

Adjusting the Respiration Amplitude (Impedance Variation)

The MPS450 offers four preconfigured settings (ohms) for respiration amplitude: 0.2,

0.5, 1.0, or 3.0.

Action in the Menu-Control Mode

1. Press the top-menu key labeled RESP. If the respiration-amplitude setting has not

been changed, the screen displays the default setting (1.0 OHMS).

2. To adjust the respiration amplitude, select AMPL repeatedly to scroll to the desired

setting. The amplitude identified on the LCD screen is active when displayed and

remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-

amplitude setting according to Table 3-4, and select RUN:

Table 3-4. Numeric Codes for Respiration-Amplitude Settings

Respiration Amplitude Setting

Numeric Code

0.2 ohms

195

0.5 ohms

196

1.0 ohms

197

3.0 ohms

198