Setting r-wave width – Fluke Biomedical MPS450 User Manual

Page 57

Advertising
background image

Cardiac Functions

ECG Testing

2

2-25

screen is active when displayed and remains active until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for R-wave rate

according to Table 2-19, and select RUN:

Table 2-19. Numeric Codes for R-Wave-Rate Settings

R-Wave-Rate Setting

Numeric Code

30 BPM

312

60 BPM

313

80 BPM

314

120 BPM

315

200 BPM

316

250 BPM

317

Setting R-Wave Width

The MPS450 offers twenty-two preconfigured settings (ms) for R-wave width: nineteen

settings ranging from 20 to 200 ms (10 ms steps), plus three settings at 8, 10, or 12 ms.

Action in the Menu-Control Mode

1. Press the top-menu key labeled PERF.
2. Select RWDET.
3. Select SEL to toggle to the screen for adjusting R-wave width (WIDTH/ampl), with

WIDTH in upper-case letters.

4. Select DOWN or UP to scroll to the desired R-wave width.
5. The width identified on the LCD screen is active when displayed and remains active

until the setting is changed.

Alternatively, in the Numeric-Control Mode, press the number keys for R-wave width

according to Table 2-20, and select RUN:

Table 2-20. Numeric Code R-Wave Width Settings

R-Wave-Width Setting

Numeric Code

8 ms

318

10 ms

319

12 ms

320

20 ms

321

30 ms

322

40 ms

323

50 ms

324

60 ms

325

70 ms

326

80 ms

327

90 ms

328