Cardiac Functions

Fetal / Maternal ECG (Option)

2

2-41

Action in the Menu-Control Mode

1. Press the top-menu key labeled MORE.
2. Select FETAL MATR.
3. Select PER to scroll to a different contraction-period setting.
4. Select IUP to scroll to the desired wave simulation: EARLY DEC (early

deceleration); LATE DEC (late deceleration); or ACC (acceleration).

5. Select RUN. On the LCD screen, RUNNING flashes while the simulation is active.
6. A manually generated IUP ends after one contraction. To repeat the simulation, select

RUN again.

7. Periodically generated IUPs repeat at the selected period. To terminate a periodic-

IUP simulation, select END.

Alternatively, in the Numeric-Control Mode, press the number keys for an IUP-wave

simulation using one of the codes in Table 2-33, and select RUN. (The selected IUP is

generated according to the specific period that was last selected, whether manual, or once

every 2, 3, or 5 minutes.)

Table 2-33. Numeric Codes for IUP Wave Settings

IUP Wave Setting

Numeric Code

IUP OFF

400

IUP EDEC (early deceleration)

402

IUP LDEC (late deceleration)

403

IUP UACC (acceleration)

404


To select a contraction period in the Numeric-Control Mode, press the number keys for

one of the settings found in Table 2-34, and select RUN. (The selected period is applied

to a run of the specific IUP-wave simulation that was last selected, whether early

deceleration, late deceleration, or acceleration.)

Table 2-34. Numeric Codes for IUP Contraction Period Settings

IUP Contraction Period Setting

Numeric Code

IUP MANUAL

405

IUP 2 MIN

406

IUP 3 MIN

407

IUP 5 MIN

408