Appendix, A. troubleshooting, Introduction – Fluke Biomedical MPS450 User Manual

Page 107: Troubleshooting, Appendix a

Advertising
background image

A-1

Appendix A

Troubleshooting

Introduction

Table A-1 lists some possible problems and their solution for repair.

Table A-1. Problem vs. Solution

Problem Probable

Cause

Solution

ECG

Monitor filters

There will be a slight loss of amplitude when
simulating NSR. Use a pulse wave to verify
amplitude more accurately.

Lead setting on monitor
incorrect

Check monitor lead select switch. Set to
Lead II.

Lead II amplitude not correct

Leads not connected
properly

Check color-coded ECG jacks.

Blood Pressure

MPS450 transducer
sensitivity not set
correctly

Check monitor manufacturer’s BP-sensitivity
requirements. Reset the MPS450 either to
5

μ

V/V/mmHg or to 40

μ

V/V/mmHg.

Wrong BP cable

Many manufacturers use plugs and
configurations with wiring differences. Check
the MPS450 BP-wiring diagram and match
with monitor specification.

Monitor not zeroed

Zero the MPS450 BP channel. Next, zero the
monitor.

Monitor sensitivity
(scaling) not correct

Adjust scaling on monitor, near range
simulated (i.e., 120 mmHg).

No BP or wrong output

Performance waveform
currently output (which
turns off BP)

Select normal physiological waveform.

Advertising