Operation – Hypertherm HT4001 Plasma Arc Cutting System User Manual

Page 105

Advertising
background image

OPERATION

6-98

5-14

HT4001

Instruction Manual

Note:

If using the H-401 power supply as a slave, note that the current control adjust knob on the
front panel of the H-401 will have no effect on setting current for the slave.

11. Set the S2 toggle switch to Test Preflow. Set the preflow flow rate(s) on the O

2

and/or N

2

flow meters using the

MV2 and/or MV3 metering valve(s). Select the test preflow rates from the Cut Charts.

Note:

If the power supply has been off for more than one hour, or if you have changed
consumable parts or have changed gases, purge gas lines by leaving system in
TEST PREFLOW for one minute.

12. Set S2 to Test Cut Flow. Set the cut flow rate on the O

2

or N

2

flow meters using the MV1 metering valve

(momentary switch). Select the test cut flow rates from the Cut Charts.

Note:

If the power supply has been off for more than one hour, or if you have changed
consumable parts or have changed gases, purge gas lines by leaving system in
TEST PREFLOW for one minute.

13. Check the water flow rate at the water flow meter and adjust with MV4 metering valve, if necessary. See the

Cut Charts for water flow rates.

14. Verify that there is a uniform conical water pattern at the front of the torch. If the pattern is irregular, shut the

power supply down at the main disconnect switch and check the nozzle and swirl ring. Replace with new parts,
if worn or damaged.

15. Set S2 to Run. The system is now ready for operation.

N

2

Flow

meter

S1

MV2 MV1

MV3

S2

MV4

O

2

Flow

meter

N

2

Press.

Gauge

Water Flow
meter

O

2

Press.

Gauge

Figure 5-9

Gas Console Reference – Daily Startup

Advertising