Amprobe – Amprobe Multitest-2000 Continuity-Tester User Manual
Page 100
AMPROBE
MULTITEST2000
8.5.2. “METER"
mode
In this mode the instrument shows the below screens according to the settings made as
per paragraph 8.1.
27.09.00 17:35:12
SINGLE PHASE
CURRENT
I1 = 30.21 A
Ipk1 = 49.53 A
ThdI = 23.06 %
freq = 50.0 Hz
CLAMP TYPE: FLEX
HARM.
WAVE
PG- PG+
Example of screen
The symbols used are described in Tab. 2.
For eventual messages displayed see appendix 1 – MESSAGES DISPLAYED.
Following keys are enabled:
F1:
to pass to "HARMONIC" mode (see paragraph 8.5.3).
F2:
to pass to "WAVE" mode (see paragraph 8.5.4).
F3/F4:
to pass to previous/next function respectively.
SAVE:
to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and
current present on the instrument inputs. This function is disabled
during a recording.