Amprobe – Amprobe Multitest-2000 Continuity-Tester User Manual
Page 27
AMPROBE
MULTITEST2000
TEST LEADS
Before each measurement always assure that the
calibration is for the cables in use. During a
continuity test, if the resistance value free of
calibration (that is the resistance value less the
calibration offset value) is negative, the symbol
ν
is displayed. Probably the calibration resistance
value stored in the instrument memory is not for
the cables in use, therefore a new calibration must
be performed.
6.1.1.1. Procedure to reset test leads calibration parameters
LOW
Ω 05.06.01
Message
>99.9
Ω:
means that the
instrument detected a
resistance higher
than 5
Ω therefore it
will proceed with
Reset procedure.
ν
>
99.9
Ω
R+ R-
----
Ω ----Ω
---mA ---mA
AUTO 0.11
Ω
FUNC
CAL
To cancel calibration
parameters it is
necessary to perform a
calibration procedure
with a resistance of
test leads higher than
5
Ω (for example with
open test leads). When
a cancellation is effected
the screen to the right
is displayed temporarily.