Before each measurement always assure that the
calibration is for the cables in use. During a
continuity test, if the resistance value free of
calibration (that is the resistance value less the
calibration offset value) is negative, the symbol

ν

is displayed. Probably the calibration resistance
value stored in the instrument memory is not for
the cables in use, therefore a new calibration must
be performed.


6.1.1.1. Procedure to reset test leads calibration parameters

LOW

Ω 05.06.01

Message

>99.9

Ω:

means that the
instrument detected a
resistance higher
than 5

Ω therefore it

will proceed with
Reset procedure.



ν

>

99.9


R+ R-
----

Ω ----Ω

---mA ---mA


AUTO 0.11

FUNC

CAL

To cancel calibration
parameters it is
necessary to perform a
calibration procedure
with a resistance of
test leads higher than
5

Ω (for example with

open test leads). When
a cancellation is effected
the screen to the right
is displayed temporarily.