The chosen configuration is the following (defined for ANALYZER function)

ANALYZER CONFIG:

Frequency: 50Hz
Transforming ratio of voltmetric transformers:

1

Type of electrical equipment:

SINGLE

Clamp Type:

FLEX

Password: enabled

RECORDER CONFIG:

Start:

Manual (the recording is started

1 minute after pressing

the

START/STOP

key)

Stop: Manual
Integration period:

15min

Recording of harmonics:

ON

Recording of Voltage anomalies (voltage Sag and Surge) :

ON

Voltage Reference for Sag and Surge detection:

230V

Upper Limit for Sag and Surge detection:

6%

Lower Limit for Sag and Surge detection:

10%

Selected voltages:

V1

Selected voltage harmonics:

THD, 01, 03, 05, 07

Selected currents:

I1

Selected current harmonics:

THD, 01, 03, 05, 07

CO-GENERATION: OFF
Powers, Pf and cos

ϕ selected:

P1

Q1i
Q1c
S1
Pf1
dpf1
Energies: Ea1
Eri1
Erc1

If the user changed the instrument’s settings can quickly resume the above configuration
using the RESET option (see paragraph 5.5).