Amprobe – Amprobe Multitest-2000 Continuity-Tester User Manual
Page 112
AMPROBE
MULTITEST2000
8.7.2. "METER"
mode
In this mode the instrument shows the below screens according to the settings made as
per paragraph 8.1.
27.09.00 17:35:12
ENERGY
SINGLE PHASE
Ea1 = 0.000 kWh
Erc1 = 0.000 kVARh
Eri1 = 0.000 kVARh
P1 = 36.38 kW
Q1 = 6.375 kVAR
S1 = 36.94 kVA
dpf1 = 0.98 i
Meas Time: 00:00:00
Meas
PG- PG+
Example of screen
The symbols used are described in Tab. 4.
For eventual messages displayed see appendix 1 – MESSAGES DISPLAYED.
Following keys are enabled:
F2:
to start / stop immediately a direct energy measurement. The
energy counters will start increasing proportionally to the active
power absorbed by the load.
The results obtained cannot be memorised.
If the active power is negative the counters will not increase.
F3/F4:
to pass to previous/next function respectively.
SAVE:
to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and
current present on the instrument inputs. This function is disabled
during a recording.