Amprobe, Multitest2000 – Amprobe Multitest-2000 Continuity-Tester User Manual
Page 137
AMPROBE
MULTITEST2000
Harmonics values are null under fixed threshold:
- DC: its values is null if it is < 2%of Fundamental or is <2% of Full Scale clamp
- 1
st
Current Harmonic: its values is null if it is < 0.2% Full Scale clamp
- 2
nd
÷ 49
th
: its values is null if it is < 2% of fundamental or is <2% of Full Scale clamp
ENVIRONMENTAL PARAMETER MEASUREMENT
Range
Accuracy
Resolution
-20°C –80 °C
0.1 °C
0
÷ 100%HR
0.1%HR
0.001Lux
÷ 20.00 Lux
0.001
÷ 0.02 Lux
0.1Lux
÷ 2000 Lux
0.1
÷ 2 Lux
1Lux
÷ 20 kLux
±(2% Reading + 2dgt)
1
÷ 20 Lux
LEAKAGE CURRENT MEASUREMENT
Range (*)
Resolution [mA]
Accuracy Input
Impedance
Overload
Protection
0.5
÷ 999.9mA
0.1mA
±(5% Reading + 2digit)
200k
Ω 5V
(*): During a recording the instrument detect only Current > 5mA with Resolution 1mA