Amprobe – Amprobe Multitest-2000 Continuity-Tester User Manual

Page 40

Advertising
background image

AMPROBE

MULTITEST2000

6.2.3. Results of "TMR" mode

M

Ω 05.06.01

Insulation Resistance

Voltage during the Test

Duration of the Test

Test mode



1.07

M



514V 15s


TMR 500V TIME:60s

FUNC

VNOM

At the end of the test if
the Insulation
resistance is lower
than R

MAX

(see Table2)

and the instrument
generated the
Nominal test Voltage
,
the instrument emits a
double beep
indicating the positive
outcome of the test

and displays a screen
similar to the screen to
the right.

The displayed result can be stored by pressing the SAVE key twice (according to

paragraph 9.1).


M

Ω 05.06.01

Maximum resistance value
that can be measured
(999

Ω is displayed if a

rated voltage of 500V was
selected, see Table2).

If the Insulation
resistance is higher
than R

MAX

(see Table2),

the instrument emits a
double beep at the end
of the test indicating



> 999

M

The symbol ">" means that

Advertising