Too many records

The quantity of recorded data + Smp exceeds the

maximum allowed (35)

Cancel some recordings after transferring

them to a PC

No Unit selected

ERR: SEQ

The Phase Sequence isn’t correct.

Check the Phase Sequence connection.

ERR: P-

The active powers shown on the right side of the message

are negative

If there isn’t a situation of co-generation

check if the clamps are properly connected

ERR: SEQ & P-

The active powers shown on the right side of the message

are negative and the Phase Sequence isn’t correct.

If there isn’t a situation of co-generation

check if the clamps are properly connected /

check the Phase Sequence connection.

ERR: CONNECTION

The instrument has detected a wrong connection to

Voltage inputs

Check the Voltage connections

Error Vref

The user set a Voltage reference not compatible with

voltage at instrument’s input.

Check Voltage Reference set in “CONFIG

RECORDER”

ERR: SYNC

The System Frequency is out of range

Check the System Frequency, check setting

in ANALYZER CONFIG.

Selection Error

There is a mismatch between the Parameter enabled and

the parameter selected for an AUX recording.

Check the parameter enabled in AUX

position and the selected parameter for

recording.

Error1

÷ Error 5

The instrument memory is damaged.

Contact HT Italia assistance