Amprobe, Recordings, Caution – Amprobe Multitest-2000 Continuity-Tester User Manual

Page 119

Advertising
background image

AMPROBE

MULTITEST2000

10. RECORDINGS

10.1. START A RECORDING

The recording function is available for ANALYZER and AUX rotary switch position.
As you can read in the paragraphs 7.2.1 and 8.2, a recording can be started manually or
automatically. Therefore, after setting all the parameters and leaving the Menu, the
instrument will start to record:

MANUALLY:

the recording will start when Instrument time reach the “00” seconds

value after pressing START/STOP.

AUTOMATICALLY:

If the operator has pressed START/STOP the instrument will

remain in stand-by until the date and time previously set, then the recording will start.
While if the operator doesn’t press START/STOP the recording will never start.


CAUTION

For recordings we recommend to use the external power supply
adapter
(optional code A0050) even the instrument allows the operator to
perform a recording using internal batteries.


If you press Start a recording without the external power supply adapter (optional code
A0050) the instrument will display a warning message “No ext supply”. Press START key
again to run the recording or press ESC to quit.


If during a recording the external power supply adapter (optional code A0050) is de-
energised, the instrument will continue the recording using the internal battery power until
the batteries are exhausted (the data stored until the definitive turning off won’t get lost).
For this we recommend you ALWAYS insert a new set of batteries before a long
recording
.

The instrument uses sophisticated algorithms to prolong the battery life. Particularly:

The instrument switches OFF the backlight Automatically after 5 seconds.