14.2. STANDARDS

14.2.1. General
Safety

EN 61010-1 + A2 (1997)

Protection classification

Class 2 - Double Insulation

Pollution degree

2

Degree of Protection:

IP50

Over-Voltage Category

CAT II 600V~ / 350V~ (phase –earth)

CAT III 600V~ / 300V~ (phase –earth)

Usage:

Indoor; max height 2000m

EMC

EN61326-1 (1997) + A1 (1998)


The Instrument comply with European Guidelines for CE mark

14.2.2. Safety Test
LOW

Ω (200mA):

IEC 61557-4

M

Ω: IEC

61557-2

PHASE SEQUENCE:

IEC 61557-7

EARTH: IEC

61557-5


14.2.3. ANALYZER
Voltage Sag and Surge

EN50160

Alternating Current Static Watt-hour meters for Active Energy

EN61036 (CLASS 2)

Alternating Current Static VAR-hour meters for Reactive Energy

IEC1268 (CLASS 3)


14.2.4. AUX
Sound Level Measurement (with External Probe):

EN 60651:1994/A1 type 1

EN 60804:1994/A2 type 1