16.7.1. Conventions on powers and power factors

As for the recognition of the type of reactive power, of the type of power factor and of the
direction of the active power, the below conventions must be applied. The stated angles
are those of phase-displacement of the current compared to the voltage (for example, in
the first panel the current is in advance from 0° to 90° compared to the voltage):

Equipment under test = Inductive Generator

Equipment under test = Capacitive Load

90°






180°

P+ = 0
Pfc+ = -1
Pfi+ = -1

Qc+ = 0
Qi+ = 0

P - = P

Pfc - = -1

Pfi - = Pf

Qc- = 0

Qi - = Q


P+ = P
Pfc+ = Pf
Pfi+ = -1
Qc+ = Q
Qi+ = 0

P - = 0
Pfc - = -1
Pfi - = -1
Qc- = 0
Qi - = 0













P+ = 0
Pfc+ = -1
Pfi+ = -1

Qc+ = 0
Qi+ = 0

P - = P
Pfc - = Pf

Pfi - = -1

Qc- = Q

Qi - = 0


P+ = P
Pfc+ = -1
Pfi+ = Pf
Qc+ = 0
Qi+ = Q

P - = 0
Pfc - = -1
Pfi - = -1
Qc- = 0
Qi - = 0







270°

Equipment under test = Capacitive Generator

Equipment under test = Inductive Load


where:

Symbol Significance

Remarks

P+

Value of the active power +

Pfc+

Capacitive power factor +

Pfi+

Inductive power factor +

Qc+

Value of the capacitive reactive power +

Qi+

Value of the inductive reactive power +

Positive parameter

(user)

P-

Value of the active power -

Pfc-

Capacitive power factor -

Pfi-

Inductive power factor -

Negative parameter

(generator)