Amprobe, Attention – Amprobe Multitest-2000 Continuity-Tester User Manual

Page 25

Advertising
background image

AMPROBE

MULTITEST2000

6.1.1. Calibrating the test leads ("CAL" Mode)

1. Connect the black and blue test leads to B1 and B4 input terminals

respectively.

B2

B3

B4

B1

Connection of instrument terminals during calibration procedure.

2. If the test leads supplied with the instrument are not long enough for the

measurement you can extend the blue cable.

3. Short-circuit the measuring cable ends making sure that the conductive parts of the

alligator clips make good contact with each other (see previous picture).

4. Press the F2 key. The instrument performs the calibration.


ν

ATTENTION:

Never disconnect the test leads when the message
"MEASURING" is displayed.

LOW

Ω 05.06.01