Chapter 5 _________________________________________________________________ Operation

WMT700 Data Message 24

WMT700 Data Message 24 reports wind measurement and self-
diagnostics data. The checksum is included in the message.
WMT700 calculates the checksum by applying 8-bit XOR for all
bytes between checksum calculation start point and end point.
The result is printed as a 2-digit hexadecimal value. The message
is in the format described below:

\ss$\ws,\wd,\wp,\wm,\Ts,\vh,\vi,\ta,\er,\se\sp\cr\lf

where

\ss

=

Checksum calculation start point, not shown in the message

$

=

Fixed text

\ws

=

Wind speed, average

\wd

=

Wind direction, average

\wp

=

Wind speed, maximum

\wm

=

Wind speed, minimum

\Ts

=

Sonic temperature

\vh

=

Heater voltage

\vi

=

Supply voltage

\ta

=

Transducer temperature

\er

=

Status code. The code is a decimal number. Each bit
corresponds to a status flag. For a list of the bits, see section
Status Flags on page 123

\se

=

Checksum calculation end point, not shown in the message

\sp

=

Print checksum (to verify message integrity)

\cr

=

CR (carriage return)

\lf

=

LF (line feed)

Example:

$03.45,76.03,03.58,03.37,21.97,23.8,23.6,23.8,0,D4<cr

><lf>

Interpretation of the example message:

- Wind speed, average: 3.45 m/s

- Wind direction, average: 76.03°

- Wind speed, maximum: 3.58 m/s

- Wind speed, minimum: 3.37 m/s

- Sonic temperature: 21.97 °C

- Heater voltage: 23.8 V

- Supply voltage: 23.6 V

- Transducer temperature: 23.8 °C

- Status code: 0

- Checksum: D4

VAISALA _______________________________________________________________________ 131