Appendix D ___________________________________________________ Configuration Parameters
APPENDIX D
CONFIGURATION PARAMETERS
This appendix lists the WMT700 configuration parameters.
Table 66
Parameter Descriptions
Parameter
Name
Default
Value
Allowed Values
Units
Description
address
A
String with a
maximum of 40
characters.
Address for WMT700. Note that the
SDI-12, ASCII, NMEA Extended, ASOS,
and MES12 profiles use the first
character only.
SDI-12 only uses digits 1 … 9.
aout1err
aout2err
1000
0 … 32000
V, A,
Hz, %
Value for AOUT1 and AOUT2 analog
output if wind measurement fails.
aout1_g
aout2_g
1
0 ... 100
Gain for AOUT1 (wind speed) and
AOUT2 (wind direction).
aout1_o
aout2_o
0
-10000 … 10000
Offset for AOUT1 and AOUT2.
aout1maxv
aout2maxv
32000
0…32000
V, A,
Hz, %
Analog output maximum value for
AOUT1 and AOUT2. The output is fixed
to this value. The unit depends on analog
output mode.
aout1minv
aout2minv
0
0…32000
V, A,
Hz, %
Analog output minimum value for AOUT1
and AOUT2. The output is fixed to this
value. The unit depends on the analog
output mode.
aout1mode
3
0 = Current
1 = Voltage
2 = Frequency
3 = Disabled
Analog output mode for AOUT1.
aout2mode
7
4 = Current
5 = Voltage
6 = Potentiometer
7 = Disabled
Analog output mode for AOUT2.
autoInt
1
0.25 ... 1000
Resolution: 0.25
s
Automatic message interval in seconds.
This parameter affects both serial
communication and analog output.
Do not select a message interval that is
shorter than the time it takes to send a
data message.
autoPort
1
1 = COM1 port
2 = COM2 port
Serial port to which WMT700 sends
automatic data messages.
autoSend
0
0 = Automatic
messages disabled
1 … 99
Automatic data message number.
Selects the data message format for
automatic messages.
cal_date
Wind calibration date.
This is a read-only parameter.
VAISALA _______________________________________________________________________ 205